Final report of the key comparison of APMP.AUV.V-K2

Page view(s)
0
Checked on
Final report of the key comparison of APMP.AUV.V-K2
Title:
Final report of the key comparison of APMP.AUV.V-K2
Journal Title:
Metrologia
OA Status:
closed
Keywords:
Publication Date:
20 June 2017
Citation:
Takashi Usuda et al 2017 Metrologia 54 09004
Abstract:
A key comparison of vibration acceleration APMP.AUV.V-K2, which is an upgrade from APMP.AUV.V-K1.1, has been made within the Asia Pacific Metrology Programme (APMP) to include three national laboratories; the CMS-ITRI (Chinese Taipei), the NIMT (Thailand) and the A*Star (Singapore). One pilot laboratory, the NMIJ (Japan), participated to link the RMO APMP results to the CIPM CC key comparison (CCAUV.V-K2), according to the decision of the CCAUV10/D7 during the 10 th CCAUV meeting. The admissible acceleration amplitude ranges from 10 m/s 2 to 200 m/s 2 over the frequency range from 40 Hz to 5 kHz. The RMO APMP results demonstrate the agreement with the key comparison reference value of CCAUV.V-K2 within the expanded uncertainties considering the armature effect of vibration exciter in a high-frequency range.
License type:
PublisherCopyrights
Funding Info:
A*STAR
Description:
This is an author-created, un-copyedited version of an article accepted for publication/published in Metrologia. IOP Publishing Ltd is not responsible for any errors or omissions in this version of the manuscript or any version derived from it. The Version of Record is available online at https://doi.org/10.1088/0026-1394/54/1A/09004
ISSN:
0026-1394
1681-7575
Files uploaded:

File Size Format Action
apmpauvv-k2.pdf 793.69 KB PDF Open